characteristic TDS3000C series electrical characteristics  
         
          TDS3012CTDS3014CTDS3032CTDS3034CTDS3052CTDS3054C
            
            
             
             | bandwidth |  
             100 MHz |  
             100 MHz |  
             300 MHz |  
             300 MHz |  
             500 MHz |  
             500 MHz |  
              
             
             | Calculated rise time (typical value) |  
             3.5 ns |  
             3.5 ns |  
             1.2 ns |  
             1.2 ns |  
             0.7 ns |  
             0.7 ns |  
              
             
             | Number of channels |  
             two |  
             four |  
             two |  
             four |  
             two |  
             four |  
              
             
             | External trigger input |  
             All models include |  
              
             
             | Sample rate per channel |  
             1.25 GS/s |  
             1.25 GS/s |  
             2.5 GS/s |  
             2.5 GS/s |  
             5 GS/s |  
             5 GS/s |  
              
             
             | Large record length |  
             10 k sample points |  
              
             
             | Vertical resolution |  
             9 bits |  
              
             
             | Vertical sensitivity, 1 M Ω |  
             1 mV/div to 10 V/div |  
              
             
             | Vertical sensitivity, 50 Ω |  
             1 mV/div to 1 V/div |  
              
             
             | Input coupling |  
             AC, DC, GND |  
              
             
             | Input impedance |  
             1 M Ω when 13 pF or 50 Ω terminals are connected in parallel |  
              
             
             | DC gain accuracy |  
             2% |  
              
             
             | Large input voltage (1 M Ω) |  
             150 VRMS, peak value ≤ 400 V |  
              
             
             | Large input voltage (50 Ω) |  
             5 VRMS, peak value ≤ 30 V |  
              
             
             | Positioning range |  
             5 div |  
              
             
             | Bandwidth limit |  
             20 MHz |  
             20 MHz |  
             20 MHz, 150 MHz |  
             20 MHz, 150 MHz |  
             20 MHz, 150 MHz |  
             20 MHz, 150 MHz |  
              
             
             | Time base range |  
             4 ns to 10 s |  
             4 ns to 10 s |  
             2 ns to 10 s |  
             2 ns to 10 s |  
             1 ns to 10 s |  
             1 ns to 10 s |  
              
             
             | Time base precision |  
             At any 1 ms interval± 20 ppm |  
              
             
             | Input/output interface |  
              
             
             | Ethernet port |  
             RJ-45 connector, supporting 10Base-T LAN |  
              
             
             | USB port |  
             Front panel USB 2.0 host port, supporting USB flash memory |  
              
             
             | GPIB port |  
             Comprehensive call/answer mode, setting and measurement (optional for TDS3GV communication module) |  
              
             
             | RS-232-C port |  
             DB-9 male connector, comprehensive sending/receiving mode; Control all modes, setting and measuring baud rate 38400 (optional for TDS3GV communication module) |  
              
             
             | VGA video port |  
             DB-15 female connector, monitor output, direct display of information on a large monitor equipped with VGA (TDS3GV communication module optional) |  
              
             
             | External trigger input |  
             BNC connector, input impedance>1 M Ω, parallel current 17 pF; Large input voltage 150 VRMS |  
              
            
            
           
         
          Acquisition mode
            
            
             
             | DPO |  
             Capture and display complex waveforms, random events, and subtle patterns in actual signal behavior. DPO can display, store and analyze the three-dimensional information of signals in real time, namely amplitude, time and amplitude distribution in each period. |  
              
             
             | Peak detection |  
             Capture high frequency and random burrs. Burrs as narrow as 1 ns can be captured. |  
              
             
             | WaveAlert® |  
             Monitor the input signal on all channels and warn the user when any waveform deviates from the captured normal waveform occurs. |  
              
             
             | sampling |  
             Data sampling only. |  
              
             
             | average value |  
             Average 2 to 512 (optional) captured waveform data |  
              
             
             | envelope |  
             The large/small value obtained on one or more acquisition signals. |  
              
             
             | Single Sequence |  
             Use the Single Sequence button to capture a single triggered capture sequence at a time. |  
              
            
            
           
         
          Trigger system
            
            
             
             | Main triggering methods |  
             Automatic triggering (supports 40ms/lattice and slower scrolling mode), normal triggering, and single sequence triggering. |  
              
             
             | B trigger |  
             Triggered after a time or event. |  
              
             
             | Time trigger range |  
             13.2 ns to 50 s |  
              
             
             | Event trigger range |  
             1 to 9999999 events. |  
              
            
            
           
         
          Trigger Type
            
            
             
             | edge |  
             Conventional level driven trigger. Positive and negative slopes on any channel. Coupling selection: DC, noise suppression, high-frequency suppression, low-frequency suppression. |  
              
             
             | video |  
             It can be triggered on all lines or single lines of NTSC, PAL, SECAM, odd/even or all fields. |  
              
             
             | Extend Video |  
             Can be triggered on broadcast and non broadcast (custom) standard specific lines and on analog high-definition TV formats (1080i, 1080p, 720p, 480p). The TDS3VID or TDS3SDI application module is required. |  
              
             
             | Pulse width (or burr) |  
             It can be triggered on a pulse width less than, greater than, equal to or not equal to the optional time limit range (39.6 ns - 50 s). |  
              
             
             | Underamplitude pulse |  
             It can be triggered on a pulse passing through a threshold, but cannot pass through the second threshold before passing through the threshold again. |  
              
             
             | Conversion rate |  
             It can be triggered at a pulse edge rate faster or slower than the set rate. It can be a rising edge, a falling edge, or both. |  
              
             
             | Code type |  
             Specify AND, OR, NAND, NOR when it is true or false at a specific time. |  
              
             
             | state |  
             Any logical state. The rising or falling edge of the clock can be triggered. You can use logic triggers on a combination of two inputs instead of four. |  
              
             
             | signal communication |  
             Provide the required isolation pulse trigger function for DS1/DS3 communication template test according to ANSI T1.102 standard. TDS3TMT application module is required. |  
              
             
             | alternate |  
             Use each active channel as the trigger source in order. |  
              
            
            
           
         
          measurement system
            
            
             
             | cursor |  
             Range, time. |  
              
             
             | Automatic waveform measurement |  
             -Display the waveform measurement values of four arbitrary combinations. Or use the measurement snapshot function to display all measurement data. Measurement items include: cycle, frequency, width, - width, rise time, fall time, duty cycle, - duty cycle, overshoot, - overshoot, high, low, large, small, peak to peak, amplitude, average, average cycle, root mean square, root mean square of cycle, burst width, delay, phase, area * 1, cycle area * 1 |  
              
             
             | Measurement statistics |  
             Average, small, large, standard deviation. The TDS3AAM application module is required. |  
              
             
             | threshold |  
             User defined automatic measurement threshold, which can be set by percentage or voltage. |  
              
             
             | Strobe |  
             You can use the screen or vertical cursor to isolate specific events that occur inside the collection for measurement. |  
              
            
            
           
         
          Waveform mathematical operation
            
            
             
             | Algebraic operation |  
             Add, subtract, multiply, divide waveforms. |  
              
             
             | FFT |  
             Spectrum amplitude. Set the FFT vertical scale to linear RMS or dBV RMS, and set the FFT window to Rectangular, Hammering, Hanning, or Blackman Harris. |  
              
             
             | Advanced mathematical operation * 1 |  
               Integral, differential, and widely defined algebraic expressions, including analog waveforms, mathematical functions, scalars, two more user adjustable variables, and parameter measurement results.  Example: (Intg (Ch1 Mean (Ch1)) x 1.414 x VAR1).   |  
              
            
            
           
         
          Waveform processing
            
            
             
             | Automatic setting |  
             On the vertical, horizontal and trigger systems, one key can automatically set all channels, and the automatic setting function can be cancelled. |  
              
             
             | Offset correction |  
             The offset correction value between channels (± 10ns) can be input manually to obtain better timing measurement value and more accurate mathematical waveform. |  
              
            
            
           
         
          Display characteristics
            
            
             
             | Display Type |  
             6.5 inch (165.1 mm) LCD TFT color display. |  
              
             
             | Display resolution |  
             640 pixels horizontal x 480 pixels vertical (VGA). |  
              
             
             | Interpolation |  
             Sin(x)/x |  
              
             
             | Waveform style |  
             Point, vector, variable afterglow, infinite afterglow. |  
              
             
             | Markings |  
             Full standard, grid, cross line, frame, NTSC, PAL, SECAM and oscilloscope (100% and 75% color bar vectors) require TDS3VID and TDS3SDI video application modules. |  
              
             
             | format |  
             YT, XY and gated XYZ (XY with Z-axis blank display is only available on 4-channel instruments). |  
              
            
            
           
         
          Power Supply
            
            
             
             | AC power supply |  
              |  
              
             
             | supply voltage |  
             100 VRMS- 240 VRMS± 10%。 |  
              
             
             | Power frequency |  
               45 Hz - 440 Hz at 100 V-120 V;  45 Hz - 66 Hz at 120 V - 240 V.   |  
              
             
             | energy consumption |  
             Large 75 W. |  
              
             
             | Power supply |  
             Requires TDS3BATC rechargeable lithium battery |  
              
             
             | Typical working hours |  
             3 hours |  
              
            
            
           
         
          Environment and safety work Non work
            
            
             
             | temperature |  
             0 º C to 50 º C |  
             -40 º C to 71 º C |  
              
             
             | humidity |  
               Working and non working humidity: when the temperature is less than or equal to 30 ℃, the relative humidity is 95% higher;  Working and non working humidity: when the temperature is less than or equal to 50 ℃, the relative humidity is 45% higher.   |  
              
             
             | height |  
             3000m |  
             15000m |  
              
             
             | electromagnetic compatibility |  
             Meet or exceed EN55011A radiation and conducted radiation standards; EN6100-3-2 AC power line harmonic radiation; EN6100-3-3 Voltage variation, fluctuation and flicker; FCC 47CFR, Part 15, Subpart B, Class A equipment; Australian EMC outline provisions. |  
              
             
             | security |  
             UL61010B-1, CSA1010.1, IEC61010-1, EN61010-1 |  
              
            
            
           
         
          Physical characteristics
            
            
             
             | instrument |  
              
             
             | External dimensions |  
             millimeter |  
             inch |  
              
             
             | width |  
             three hundred and seventy-five |  
             fourteen point eight |  
              
             
             | height |  
             one hundred and seventy-six |  
             six point nine |  
              
             
             | thickness |  
             one hundred and forty-nine |  
             five point nine |  
              
             
             | weight |  
             kg . |  
             pound |  
              
             
             | Net weight of instrument |  
             three point two |  
             seven |  
              
             
             | Install the battery |  
             four point five |  
             nine point eight |  
              
             
             | Instrument delivery |  
              
             
             | Packaging appearance dimension |  
             millimeter |  
             inch |  
              
             
             | width |  
             five hundred and two |  
             nineteen point eight |  
              
             
             | height |  
             three hundred and seventy-five |  
             fourteen point eight |  
              
             
             | thickness |  
             three hundred and sixty-nine |  
             fourteen point five |  
              
             
             | Rack mount (RM3000) |  
              
             
             | External dimensions |  
             millimeter |  
             inch |  
              
             
             | width |  
             four hundred and eighty-four |  
             nineteen |  
              
             
             | height |  
             one hundred and seventy-eight |  
             seven |  
              
             
             | thickness |  
             one hundred and fifty-two |  
             six |  
              
            
            
           *1 TDS3AAM module is required.   |